800-444-3940

AHP300™ Maintenance

Required Preventative Maintenance for the AHP300 Transport Ventilator

The following table denotes the preventative maintenance schedule for the AHP300 series transport ventilator. This service is available from Allied Healthcare Products Inc. or through authorized AHP300 service centers. Note that some of the service below can be performed by the end user provided that they have access to the appropriate equipment. Consult the AHP300 operation manual for a detailed list of the procedures & equipment needed at each service interval.

Year

Requirement

Notes

1

  • Yearly Verification of Calibration and Function

May be performed by Allied or in the field per the instruction manual.

2

  • Yearly Verification of Calibration and Function

May be performed by Allied or in the field per the instruction manual.

3

  • Yearly Verification of Calibration and Function
  • Battery Replacement
  • Test Unit

Must be performed by Allied Healthcare Products or by factory trained Service Center.

4

  • Yearly Verification of Calibration and Function

May be performed by Allied or in the field per the instruction manual.

5

  • Yearly Verification of Calibration and Function

May be performed by Allied or in the field per the instruction manual.

6

  • Yearly Verification of Calibration and Function
  • Battery Replacement
  • Replace Seals
  • Replace Flow Sensors
  • Replace Compressor
  • Test Unit

Must be performed by Allied Healthcare Products or by factory trained Service Center.